Norm

ASTM D3454 - 11 en

Standard Test Method for Radium-226 in Water

  • Deze norm is ingetrokken sinds 01-10-2018

50,60 61,23 Incl BTW

Over deze norm

Status Ingetrokken
Aantal pagina's 9
Gepubliceerd op 15-06-2011
Taal Engels

1.1 This test method covers the measurement of soluble, suspended, and total radium-226 in water in concentrations above 3.7 × 103 Bq/L. This test method is not applicable to the measurement of other radium isotopes.

1.2 This test method may be used for quantitative measurements by calibrating with a radium-226 standard, or for relative measurements by comparing the measurements made with each other.

1.3 This test method does not meet the current requirements of Practice D2777.

1.4 The values stated in SI units are to be regarded as the standard. The inch-pound units given in parentheses are for information only.

1.5 Hydrofluoric acid (HF) is very hazardous and should be used in a well-ventilated hood. Wear rubber gloves, safety glasses or goggles, and a laboratory coat. Avoid breathing any HF fumes. Clean up all spills promptly and wash thoroughly after using HF.

1.6 This standard does not purport to address all of the other safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.