Standard Practice for Set-up, Calibration, and Quality Control of Instruments Used for Radioactivity Measurements

78,39

Over deze norm

Status Definitief
Aantal pagina's 24
Commissie D19.04
Gepubliceerd op 01-06-2014
Taal Engels

1.1 This practice covers consensus criteria for the calibration and quality control of nuclear instruments. This practice is provided for establishing appropriate quality control parameters at instrument startup, calibration of nuclear counting instruments and the continuing monitoring of quality control parameters. Calibrations are usually performed to establish the operating parameters of the instrument. This practice addresses the typically used nuclear counting instruments: alpha spectrometer, gamma spectrometer, gas proportional counter and liquid scintillation counter.

1.2 The values stated in SI units are to be regarded as standard. The values given in parentheses are mathematical conversions that are provided for information only and are not considered standard.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Details

ICS-code 17.240
Nederlandse titel Standard Practice for Set-up, Calibration, and Quality Control of Instruments Used for Radioactivity Measurements
Engelse titel Standard Practice for Set-up, Calibration, and Quality Control of Instruments Used for Radioactivity Measurements