Norm

NEN-EN 12543-4:1997 Ontw. en

Niet-destructief onderzoek - Karakterisering van focussen (brandvlekken) in industriële röntgentoestellen voor gebruik in niet-destructief onderzoek - Deel 4: Randmethode

  • Deze norm is ingetrokken sinds 04-10-1999

16,10

Over deze norm

Status Ingetrokken
Aantal pagina's 8
Gepubliceerd op 01-02-1997
Taal Engels
The image quality and the resolution of X-ray images depend highly on the characteristics of the focal spot. The imaging qualities of a focal spot are based on the twodimensional intensity distribution in the object plane. This standard describes the measurement of focal spot dimensions above 0,3 mm of X-ray systems up to and including 500 kV tube voltage, by means of radiographs of sharp edges. The edge method is especially useful for the checking of focal spots under field conditions in order to find changes of the focal spot. It cannot be used for an absolute measurement of the focal spot. For absolute measurements the method according to Annex A shall be applied. If the focal spot size has to be certified, the measurement method shall meet the values of EN 12543-1 within an accuracy of ±10%.

Details

ICS-code 19.100
Nederlandse titel Niet-destructief onderzoek - Karakterisering van focussen (brandvlekken) in industriële röntgentoestellen voor gebruik in niet-destructief onderzoek - Deel 4: Randmethode
Engelse titel Non-destructive testing - Characteristics of focal spots in industrial X-ray tube assemblies for use in non-destructive testing - Part 4: Edge method
Vervangen door