NEN-EN-IEC 61558-2-15:2001 en;fr

Veiligheid van energietransformatoren, voedingseenheden en dergelijke - Deel 2-15: Bijzondere bepalingen voor beschermingstransformatoren ten behoeve van medisch gebruikte ruimten

  • Deze norm is ingetrokken sinds 14-02-2012

58,16

Over deze norm

Status Ingetrokken
Aantal pagina's 31
Gepubliceerd op 01-07-2001
Taal Engels, Frans
Applies to stationary, single-phase or polyphase, air-cooled (natural or forced) isolating transformers for the supply of group II medical locations, designed to be permanently connected to the fixed wiring of IT supply system, having a rated supply voltage not exceeding 1000 V a.c. and rated frequency not exceeding 500 Hz. The rated output shall not be less than 3 kVA and shall not exceed 10 kVA. The no-load output voltage and the rated output voltage shall not exceed 250 V a.c. single-phase or polyphase (phase-to-phase voltage). Isolating transformers are used where double or reinforced insulation between circuits is required by the installation rules or by the appliance specification. This standard is applicable to dry-type transformers. The winding may be encapsulated or non-encapsulated.

Details

ICS-code 29.180
Nederlandse titel Veiligheid van energietransformatoren, voedingseenheden en dergelijke - Deel 2-15: Bijzondere bepalingen voor beschermingstransformatoren ten behoeve van medisch gebruikte ruimten
Engelse titel Safety of power transformers, power supply units and similar - Part 2-15: Particular requirements for isolating transformers for the supply of medical locations